Friday, April 20, 2012

[Catalogs] AIR PURGE TYPE BALLAST & DRAFT LEVEL GAUGING SYSTEM



AIR PURGE TYPE BALLAST & DRAFT LEVEL GAUGING SYSTEM





Principle of Operation
The principle of operation is based on the measurement of the hydrostatic pressure by providing a constant low flow of air or neutral gas into a probe which opens at the tank bottom. the output pneumatic signal of the modulator is fed into P/I converter and is changed to electric signal (4~20mA) in 2 wire by P/I convertor. The electric ouptup signal (4~20mA) can be connected to C.R.T display cargo system, Digital indicator, analogue type indicators, etc. or a combination of these system.

Application
  • Liquid level of measuring depth pressure is indicated for direct reading, and then the high precision is achieved. 
  • The construction is simple and the handling and maintenance are easy.
  • 4~20mA output signal 2-wires.
  • Blowing valve is select type (change over type). 
  • Included stop valve at air purge head, other pressure in no need for line leakage test.


Features
  • Ballast tank remote reading
  • Draft remote reading
  • fuel oil tank remote reading
  • All liquids even viscous ones 

CONTACT INFO
MARSEN Co.,ltd
3153, Myungji, Gangseo gu, Busan, South Korea
Phone 82-51-831-2108
Fax 82-51-831-2109
www.marsen.co.kr

Komec Website: www.komec.kr, B2B Marketplace: www.shipsol.com
Chinese Blog : blog.sina.com.cn/komec
General Support: +82-51-972-6474, +82-51-972-6478
E-mail: gsc@komea.kr
KOMEC QR CODE / KOMEC Blog QR CODE